U.S. flag

An official website of the United States government, Department of Justice.

NCJRS Virtual Library

The Virtual Library houses over 235,000 criminal justice resources, including all known OJP works.
Click here to search the NCJRS Virtual Library

AmpFlSTR Profiler Plus Short Tandem Repeat DNA Analysis of Casework Samples, Mixture Samples, and Nonhuman DNA Samples Amplified Under Reduced PCR Volume Conditions (25 uL)

NCJ Number
203117
Journal
Journal of Forensic Sciences Volume: 48 Issue: 5 Dated: September 2003 Pages: 1014-1034
Author(s)
Chantal J. Fregeau Ph.D.; Kathy L. Bowen B.Sc.; Benoit Leclair Ph.D.; Isabelle Trudel M.Sc.; Lucy Bishop B.Sc.; Ron M. Fourney Ph.D.
Date Published
September 2003
Length
21 pages
Annotation
This paper presents the methodologies and findings of some of the studies that were conducted during the validation of the AmpFlSTR Profiler Plus STR amplification system, using reduced PCR volume conditions (25 ml).
Abstract
As part of the validation of the AmpFlSTR Profiler Plus STR system under reduced polymerase chain reaction (PCR) volume conditions, 275 casework samples were processed. As a means of improving the odds of obtaining balanced and complete profiles for casework samples that show partial profiles or profiles with a slope, several amplification conditions were involved. In addition, a series of 10 simulated mixtures was prepared to establish the limit of detection of a minor profile using the Profiler Plus multiplex system. Two amounts of total DNA (1 and 2 ng) and different ratios that ranged from 1:20 to 20:1 were selected to cover a wide range of potential casework mixtures. The specificity of the Profiler Plus multiplex amplification reaction was also examined under reduced PCR volume conditions by using a large spectrum of nonhuman DNA's, varying the amounts of template DNA (2.5 and 50 ng). All the results indicate the robustness and reliability of the AmpFlSTR Profiler Plus system when used under the experimental conditions described in this report. Reducing the PCR volume to 25 ml allows major cost saving per sample analysis while maintaining high-quality profiles. Thus, very high discrimination can be achieved with minimal analytical consumption of crucial evidentiary samples. 8 figures and 35 references